Every cleanroom answers to one number, and that number comes from a particle counter. Whether you are classifying a new room, qualifying it for GMP work, or watching it day to day, the counter tells you if the air is doing its job.
The instrument itself is simple. How it gets used is where rooms pass or fail. Sampling location, sample volume, tubing, flow rate, and timing all change the result, and most disputed counts trace back to the method rather than the room. This guide explains how cleanroom particle counting works and how to do it so the numbers hold up.
What Cleanroom Particle Counting Actually Measures
A particle counter draws a known volume of air through a sensing chamber and reports how many airborne particles of each size were in it. The result is a concentration, usually written as particles per cubic metre for a given size threshold, such as 0.5 micrometres and larger.
It counts particles, not organisms. A viable particle, like a bacterium riding on a skin flake, shows up in the count only as a particle of some size. Microbial monitoring with settle plates, contact plates, and active air samplers is a separate program that sits beside particle counting.
Particle counts are the language cleanroom standards use for air cleanliness. ISO 14644-1 sets classes by maximum particle concentration, and EU-style GMP grades in Health Canada’s Annex 1 guidance on environmental and process monitoring add limits for both at rest and in operation.
How an Optical Particle Counter Works
Most cleanroom counters are light scattering instruments. Sample air passes through a laser beam, and each particle scatters a pulse of light onto a detector. The height of the pulse tells the counter how big the particle is, and the number of pulses tells it how many there were.
The counter sorts pulses into size channels, for example 0.3, 0.5, 1.0, 5.0 micrometres and larger. A pump holds a fixed flow rate so that the volume of air sampled is known. Counts divided by that volume give concentration.
Two details matter for credibility. First, light scattering counters are calibrated against a standard, and ISO 21501-4 is the one normally cited for this type of instrument. Second, the calibration is tied to the flow rate, so a counter running with a weak pump or a kinked line is reporting numbers that no longer mean what the certificate says.

Classification Counts Versus Routine Monitoring
People use the same instrument for two different jobs, and the rules differ. Classification and requalification are formal events that prove the room meets its class at defined locations. Routine monitoring is the ongoing watch that tells you the room has not drifted since.
Classification follows the sampling plan in ISO 14644-1, and the test procedures are covered in our guide to ISO 14644-3 test methods. Routine monitoring follows your own contamination control strategy, shaped by risk and by the regulator for your product.
| Item | Classification or requalification | Routine monitoring |
|---|---|---|
| Purpose | Prove the room meets its stated class | Show the room stays in control |
| Who sets the plan | ISO 14644-1 sampling rules | Your risk assessment and quality system |
| Typical state | At rest, sometimes operational | Operational, during real work |
| Instrument | Usually a portable counter, often by a certifier | Portable or fixed remote system |
| Frequency | Set intervals under ISO 14644-2 | Per procedure, continuous for the highest grades |
| Output | Pass or fail report by location | Trends, alerts, action limits |
Related Article: Cleanroom Certification: What ISO 14644-1 Testing Actually Involves
ISO 14644-1 Class Limits at a Glance
The ISO class sets the maximum number of particles allowed in a cubic metre of air at each size. The table below shows the 0.5 micrometre limit for the classes most pharmaceutical and lab rooms use, alongside the Annex 1 grade that lines up with the same at-rest number.
| ISO class | Max particles per m3, 0.5 µm and larger | Matching at-rest value in Annex 1 table |
|---|---|---|
| ISO 5 | 3,520 | Grade A and Grade B at rest |
| ISO 6 | 35,200 | No direct grade |
| ISO 7 | 352,000 | Grade C at rest |
| ISO 8 | 3,520,000 | Grade D at rest |
Treat this as a quick reference. The standard itself defines classes across a wider range of sizes, and the current edition should be your working copy when you write a protocol.
Health Canada Annex 1 Limits by Grade
For sterile manufacturing, Health Canada’s Annex 1 guidance publishes a table of maximum permitted total particle concentration for monitoring. It gives both at rest and in operation values at 0.5 and 5.0 micrometres.
| Grade | 0.5 µm, at rest / in operation | 5.0 µm, at rest / in operation |
|---|---|---|
| A | 3,520 / 3,520 | 29 / 29 |
| B | 3,520 / 352,000 | 29 / 2,930 |
| C | 352,000 / 3,520,000 | 2,930 / 29,300 |
| D | 3,520,000 / not predetermined | 29,300 / not predetermined |
The guidance says the at-rest limits should be reached after a short clean up period, with a guidance value of under 20 minutes. For Grade D in operation, the manufacturer sets the limit from risk assessment and routine data.
The guidance also expects the Grade A area to be monitored continuously for both sizes, with a sample flow of at least 28 litres (1 cubic foot) per minute. Grade B can use a similar system, though the sample frequency may be lower.
Choosing Sampling Locations
Where you put the probe decides what the count means. For classification, ISO 14644-1 ties the minimum number of locations to the floor area of the room, so a bigger room needs more points. The locations should be spread to represent the whole area and placed at the working height.
For routine monitoring, location choice should follow risk. Put probes where product or critical surfaces are exposed, near doors and pass-throughs where air from lower grades can enter, and downstream of where people work. A point in a quiet corner that nobody could contaminate tells you very little.
A few habits make locations defensible:
- Mark every location on a drawing and give it an ID that stays the same from one test to the next.
- Keep the inlet at working height, not at the ceiling or the floor.
- Point the probe into the airflow in unidirectional zones, and keep it clear of surfaces and equipment that would disturb flow.
- Record any location you moved and why, so a trend break is not mistaken for a room problem.
Airlock placement and pressure cascades also change which points matter most. Our guides on cleanroom airlocks and on positive and negative pressure cleanrooms explain where the air travels, which is where probes belong.
Sample Volume and Sampling Time
A count is only as good as the volume behind it. Sample too little air in a clean room and you can see zero particles even when the real concentration sits near the limit. ISO 14644-1:2015 handles this with a minimum single sample volume formula tied to the class limit, with an absolute floor of a couple of litres and a minimum sampling time of about a minute per location. Check the exact wording in your copy before you hard code it into a protocol.
The practical effect is that cleaner classes need bigger samples. An ISO 5 room needs far more air per location than an ISO 8 room to give a statistically meaningful result.
Sampling time follows from volume and flow rate. One vendor guide puts a 1 cubic metre sample at about 35 minutes for a 1 cubic foot per minute counter and about 20 minutes at 50 litres per minute. That is why higher flow counters are popular for Grade A and ISO 5 work.
| Counter flow rate | Time to sample 1 m3 | Where it is usually used |
|---|---|---|
| 1 cubic foot per minute (about 28.3 L/min) | About 35 minutes | General classification and monitoring |
| 50 L/min | About 20 minutes | Cleaner rooms where large volumes are needed |
| 100 L/min | About 10 minutes | Very clean zones, subject to the counter being qualified for it |
The first two rows come from a published vendor guide. The 100 litre row is simple arithmetic from the flow rate and is a rule of thumb only.
Flow Rate, Size Channels, and Calibration
Flow rate is the part of the instrument people check least and trust most. Counting efficiency, sizing accuracy, and the sample volume calculation all assume the pump delivers the flow stated on the certificate.
Channel selection matters just as much. A counter that only reports 0.5 micrometres and larger cannot give you the 5 micrometre numbers an Annex 1 program asks for. Confirm that the smallest and largest channels cover every size in your protocol before you buy or rent a unit.
Calibration should be current, traceable, and done to a recognized standard for light scattering counters. Ask for the certificate and keep it with the test report. Many sites also run a zero count test before each session, with a filter fitted to the inlet, to show the instrument is not generating false counts.
Handheld, Portable, and Remote Counters
There are three common ways to count, and each fits a different job. A handheld or portable counter goes where the technician carries it. A fixed remote system uses probes and tubing connected to a central counter or manifold. A fixed counter with a built-in sensor sits in the room and reports continuously.
| Type | Best for | Watch out for |
|---|---|---|
| Handheld or portable | Classification, requalification, troubleshooting | Operator technique, location repeatability |
| Remote sampling with tubing | Continuous monitoring of critical zones | Tubing length and bends cause particle loss |
| Fixed in-room sensor | Trending in lower grade rooms | Sensor placement and drift between calibrations |
Neither type replaces the other. Portable counters prove classification and chase problems, and fixed systems catch events as they happen. Most pharmaceutical sites use both.
Related Article: Environmental Monitoring Systems for Controlled Environment Rooms
Probes, Tubing, and Isokinetic Sampling
Particles are lost in transit. Large particles settle or stick to tubing walls, especially around bends, so a long, curved sample line under counts the sizes that matter most for contamination. The fix is to keep tubing short, use smooth bore conductive material, and avoid sharp turns.
In unidirectional airflow, the probe inlet should match the air velocity so that the sample is representative. That is the idea behind isokinetic sampling heads. A mismatch between probe velocity and room airflow skews the count toward smaller or larger particles.
The tubing and probe configuration should be written into the monitoring procedure and treated as part of the validated setup. Changing a tube length without a documented check can invalidate the qualification of that sample point.

How to Run a Particle Count Step by Step
The sequence below is a good baseline for a portable counter during classification or requalification. Adapt it to your own SOP and the standard that governs your room.
- Confirm the counter calibration is current and the battery and pump are in good condition.
- Run a zero count check with a filter on the inlet and record the result.
- Verify the room state, at rest or operational, and that airflow and pressure readings are stable.
- Gown to the standard of the room, and bring only equipment that has been cleaned for the area.
- Set up at the first location, at working height, with the probe pointed into the airflow.
- Enter the sample volume and time, then start the run and step away from the probe so you do not disturb the air.
- Record the counts for every channel, the time, and the location ID.
- Repeat at each location, and repeat any location where a count was unusually high.
- Finish with a second zero count if your procedure requires it, then export and review the data.
Gowning discipline matters here because the person holding the counter is usually the biggest source of particles in the room. Our cleanroom gowning procedure guide covers how to keep that contribution low.
At Rest and Operational Counts
The room state changes the answer. At rest means the room is complete, equipment is installed and running, and no staff are present. Operational means the room is in use with people and processes in place.
At rest counts show what the room and HVAC can achieve on their own. Operational counts show what the people and process add. Annex 1 style programs set separate limits for each, as the table above shows, and the gap between them is often where trouble hides.
Before an at-rest test, allow the room to recover after any activity. The Health Canada guidance mentions a clean up period with a guidance value of under 20 minutes. Starting a count too early records the wake of the last shift rather than the true state of the room.
Alert Limits and Action Limits
A limit from a standard tells you when the room fails. It does not tell you when the room is starting to drift. That is the job of alert and action levels.
Health Canada’s guidance says appropriate alert levels and action limits should be set for total particle monitoring, that alarms should trigger when alert levels are exceeded, and that exceeding action limits should lead to a root cause investigation. It does not give numeric alert levels, so you set them from your own history.
A workable approach is to collect several months of data from each location, find the normal range, and set the alert level well inside the regulatory limit. The action limit then sits at or just below the regulatory number. Review the levels after any major change in the room, the HVAC, or the process.
Common Particle Counting Mistakes
Most bad counts are avoidable. These are the mistakes we see come up repeatedly in qualification and monitoring programs:
- Counting too soon after cleaning or an intervention, which records disturbance rather than the room.
- Using long or kinked tubing, which loses larger particles before they reach the sensor.
- Letting the operator stand upstream of the probe, in the path of the airflow.
- Skipping the zero count, then explaining odd readings after the fact.
- Moving a sample location without recording it, which breaks trending.
- Running a counter past its calibration date, which can void the whole test.
- Ignoring the 5 micrometre channel, which often flags problems before the 0.5 micrometre channel does.
- Setting alert levels equal to the regulatory limit, so there is no early warning.
Reading Results and Investigating Excursions
A single high count is a prompt to look, not a verdict. Start by checking the basics: was the counter in calibration, did the zero count pass, was the tubing intact, and did anything happen near the probe during the run.
If the count is real, look at the pattern. A spike at one location points to a local source such as a door, a person, or a piece of equipment. Elevated counts across the whole room point to the air side, such as a loaded or leaking filter, a low air change rate, or a pressure problem.
HEPA filter condition is the usual suspect when large areas fail. Our explanation of HEPA and ULPA filtration shows what to look at, and the guide to common cleanroom contaminants helps trace the particle source once the air side is cleared.
Bigger particles tend to point at people, materials, and cleaning. Smaller particles tend to point at the air supply and filtration. Counting in several channels gives you that clue for free.
When the Room Itself Is the Problem
Sometimes the count is right and the room really cannot hold its class. That usually comes down to design, not technique. Too few air changes, poor return placement, leaky wall or ceiling joints, and unstable pressure cascades all push counts up no matter how carefully you sample.
Cleaning also affects what you count. A well run program, like the one in our cleanroom cleaning procedure guide, keeps surface and air counts from creeping up between tests.
On the design side, cleanroom HVAC design and the construction rules in ISO 14644-4 are where class performance is won or lost. Fix the cause upstream and the particle counts take care of themselves.
Related Article: ISO 14644-2: Cleanroom Monitoring and Requalification Explained
Documentation to Keep
A particle count that is not documented might as well not have happened. Auditors will ask to see the evidence behind every number.
At a minimum, keep these records with each test:
- Counter model, serial number, and calibration certificate with expiry date
- Zero count result before and after the session
- Room ID, state at the time of the test, and airflow and pressure readings
- Sampling plan with a drawing of locations and their IDs
- Sample volume, sample time, flow rate, and tubing configuration
- Raw counts by channel and location, with operator name and date
- Any deviation, repeat test, or location change with the reason
Particle counting also feeds the wider qualification package. Our guide to IQ, OQ, and PQ validation shows where these records sit in the larger picture.
Frequently Asked Questions
What is cleanroom particle counting?
Cleanroom particle counting is the measurement of airborne particles in a known volume of air using an optical counter. Results are reported as particles per cubic metre at set size thresholds, and they are compared to the limits for the room’s ISO class or GMP grade.
How does a cleanroom particle counter work?
A pump pulls air through a laser beam inside the counter. Each particle scatters light onto a detector, and the strength of the pulse indicates size. The counter tallies pulses into size channels and divides by the sampled air volume to report concentration.
What particle sizes should I count?
Most programs report 0.5 micrometres and larger, and sterile manufacturing under Annex 1 style rules also reports 5.0 micrometres and larger. Some sites add smaller channels for research or semiconductor work. Choose channels that cover every size named in your protocol or standard.
How many sampling locations do I need?
For classification, ISO 14644-1 sets the minimum number of locations based on the room’s floor area, so bigger rooms need more points. For routine monitoring, you pick locations by risk, focusing on exposed product, critical surfaces, doors, and areas where staff work.
How much air should I sample at each location?
ISO 14644-1:2015 uses a formula tied to the class limit, so cleaner rooms need larger samples. Many Grade A and ISO 5 programs sample 1 cubic metre per location. Always confirm the exact requirement in the current edition of the standard.
How often should I count particles?
It depends on class and risk. ISO 14644-2 sets requalification intervals for classification. Annex 1 style guidance expects continuous monitoring of the Grade A area and a lower sample frequency for Grade B. Lower grades follow your risk assessment.
How often should a particle counter be calibrated?
Most manufacturers and quality systems call for calibration at least once a year, and some sites do it more often. Light scattering counters are normally calibrated to ISO 21501-4. Keep the certificate with every test report so the count stays traceable.
What is a zero count test?
A zero count test runs the counter with a filter on the inlet to confirm it is not generating false counts. The result should be zero or very close to it. Many sites record it before and after each session as a simple instrument check.
Why do long tubing runs affect counts?
Larger particles settle or stick to tubing walls, especially at bends, so a long sample line under reports the bigger sizes. Keep lines as short and straight as the design allows, use smooth conductive tubing, and document the configuration as part of the validated setup.
What is the difference between at rest and operational counts?
At rest means the room is complete and running with no personnel present. Operational means people and process are active. Limits for the operational state are looser for most grades, and the gap between the two numbers helps show how much contamination the work adds.
What should I do if a count exceeds the limit?
Check the counter calibration, the zero count, tubing, and any activity near the probe. If the count is real, look at the pattern across locations and sizes, then investigate air supply, filters, people, and cleaning. Health Canada guidance expects a root cause investigation when action limits are exceeded.
Can I use the same counter for classification and monitoring?
Often yes, if it has the right channels, flow rate, and calibration for both jobs. Many sites use portable counters for classification and troubleshooting and a fixed or remote system for continuous monitoring, because the two jobs have different needs.
Get Help With Cleanroom Qualification and Monitoring
Good counts start with a room that was built and commissioned to hold its class. Cantrol International has designed, built, and validated controlled environments from Markham, Ontario since 1989, with in-house engineering and validation across North America. If you are planning a new cleanroom or need a second opinion on a room that struggles to hold its numbers, contact our team, read how we work on our process page, browse our cleanroom solutions, or request a quote. You can also start from the Cantrol International homepage.